User's Guide ______________________________________________________________________
84 __________________________________________________________________ M210796EN-A
Using Serial Line
SERI
Use the serial line command SERI [b p d s] to set communication
settings for the user port.
SERI [b p d s]
where
b = Bit rate (110, 150, 300, 600, 1200, 2400, 4800, 9600,19200,
38400, 57600, 115200)
p = Parity (n = none, e = even, o = odd)
d = Data bits (7 or 8)
s = Stop bits (1 or 2)
Example:
>SERI 600 N 8 1
600 N 8 1
>
You need to reset the transmitter to activate the new communication
settings set with command SERI.
The settings can be changed one parameter at a time or all parameters
at once:
>SERI O changing parity only
4800 O 7 1
>SERI 600 N 8 1
changing all parameters
600 N 8 1
>
You can use the SERI command to change/view the user port settings
even if you are currently connected to the service port.
SMODE
Use the command SMODE to set the user port start-up operating
mode.
SMODE [xxxx]
where