User's Guide ______________________________________________________________________
68 __________________________________________________________________ M210796EN-A
List of Serial Commands
The bold text in the brackets is the default setting. Enter commands by
typing them on your computer and pressing the Enter key <cr>.
General commands
? Output the information about the
device
?? Output the information about the
device in POLL-state
ECHO [ON/OFF] Turn the serial interface echo
ON/OFF
FIND All devices in POLL mode are sent
their addresses
HELP List the most common commands
ERRS List present transmitter errors
SERI [baud p d s] User port settings (default: 4800 E
7 1) baud: 300…115200
LOCK Lock the menu/disable keypad
XHEAT Sensor heating
Measurement commands
R Start the continuous outputting
S Stop the continuous outputting
INTV [0…255 S/MIN/H] Set the continuous output interval
(for RUN mode)
SEND [0…99] Output the reading once
SEND D Outputting the reading with the
raw data
SCOM Define a user-specific SEND
command for one message output
SMODE [STOP/RUN/
POLL/SEND] Set the serial interface (change
operating mode)
SDELAY View or set user port (RS232 or
RS485) answer minimum delay.
ADDR [0…99] Set the transmitter address (for
POLL mode)
OPEN [0…99] Open temporarily connection to
the POLL-mode device
CLOSE Close the connection (back to
POLL mode)