Example:
RH=***.* %RH T= 31.0 'C
The output mode can be changed with the command FORM.
Outputting Reading With Raw Data
SEND D
Example:
>send d
24.1720 15.0399 -3.5743 189.2324 15.0709 15.0399
23.9765
Where the readings (from the left) are:
24.1720 = Temperature of the humidity probe ( ºC)
15.0399 = RH (%RH)
-3.5743 = Tdf (C)
189.2324 = Capasitance (pF)
15.0709 = RH raw: calculated from scaled capasitance (%RH)
15.0399 = Enchancement factor corrected RH (%RH)
23.9765 = Temperature of the additional temperature probe
(optional) (ºC)
SCOM
The SCOM command is used to define a user specific SEND
command for one message output. The standard SEND command of
the transmitter will always function normally whatever the SCOM
definition may be.
Example of setting a P command for one message output:
>scom
Send command : ? p <cr>
>
Any previous SCOM definition may be removed with the following
command:
>scom
Send command : p ? <esc>