Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 99
Example:
>itest 20 5
Ch1 (Td ) : * 20.000 mA H'672A
Ch2 (T ) : * 5.000 mA H'34F9
>itest
Ch1 (Td ) : -23.204 'C 16.238 mA H'FFFE
Ch2 (T ) : 22.889 'C 8.573 mA H'5950
>
Analog Output Fault Indication
Setting
Factory default state for analog outputs during error condition is 0 V/
0 mA. Please be careful when selecting the new error value. The error
state of the transmitter should not cause unexpected problems in
process monitoring.
Use the display/keypad to set the analog output fault indication.
1. Press any of the arrow keys to open the MAIN MENU.
2. Select Interfaces and press the right arrow key.
3. Select Analog Outputs and press the right arrow key.
4. Select Output 1/2/3 and press the right arrow key.
5. Select Fault indication. Press the SET key. Enter the fault
indication value by using the arrow keys. Press the OK key to
confirm your setting. This value is outputted if a transmitter
error occurs.
6. Press the EXIT key to return to the basic display.
AERR
Use the serial line AERR command to change the error output.
AERR
Example:
>aerr
Ch1 error out : 0.000V ? 5.0
Ch2 error out : 0.000V ? 5.0
>