Vaisala PTU300 Podręcznik Użytkownika Strona 78

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 173
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 77
User's Guide ______________________________________________________________________
76 __________________________________________________________________ M210796EN-A
Other examples:
>form "RH=" 4.2 rh U5 #t "T=" t U3 #r #n
OK
>RH= 14.98%RH T= 74.68'F
>form "Tfrost=" tdf U3 #t "Temp=" t U3 #r#n
OK
>Tfrost= 36.0'C Temp= 31.0'C
Command ‘FORM ’ will return the default output format. The default
output format depends on the device configuration.
>form /
>send
RH= 98.4 %RH T= 31.1 'C
>
UNIT
Use this command to select metric or non-metric output units. With
the command you can also set P units.
UNIT [x] [y]
where
x
= M or N or P
y
= Pressure Units (see Table 4 on page 15)
where
M = Metric units
N = Non-metric units
P = Pressure
Examples of changing the units:
>unit n
Output units : non metric
>unit m
Output units : metric
>unit p torr
P units : torr
>unit p hpa
P units : hPa
Przeglądanie stron 77
1 2 ... 73 74 75 76 77 78 79 80 81 82 83 ... 172 173

Komentarze do niniejszej Instrukcji

Brak uwag