User's Guide ______________________________________________________________________
130 _________________________________________________________________ M210796EN-A
P1 6.reading : 0.000 ? 1150
P1 6.correction: 0.000 ? 0.33
P1 7.reading : 0.000 ? 1200
P1 7.correction: 0.000 ? 0.34
P1 8.reading : 0.000 ?
P1 8.correction: 0.000 ?
MPC
Use the command MPC to view current status of the multipoint
corrections.
Example:
>mpc
P1 multi adj. : ON
P1 1.reading : 900.000
P1 1.correction: 0.200
P1 2.reading : 950.000
P1 2.correction: 0.220
P1 3.reading : 1000.000
P1 3.correction: 0.270
P1 4.reading : 1050.000
P1 4.correction: 0.310
P1 5.reading : 1100.000
P1 5.correction: 0.320
P1 6.reading : 1150.000
P1 6.correction: 0.330
P1 7.reading : 1200.000
P1 7.correction: 0.340
P1 8.reading : 0.000
P1 8.correction: 0.000
>
Relative Humidity Adjustment
Using Push-Buttons
A simple push-button adjustment is carried out by using two relative
humidity references: 11 % RH (LiCl) and 75 % RH (NaCl). The three
buttons needed for the adjustment are located on the motherboard, at
the upper left corner of the transmitter.
1. Carry out the chemical purge (if available).
LiCl reference
2. Press the ADJ button (see Figure 50 on page 126) to open the
adjustment mode. The indicator led starts flashing.