Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 83
User Port Serial Settings
Using Display/Keypad
The communication settings for the user port can be changed via the
serial line or by using the optional display/keypad. The
communication settings for the service port are fixed and not
changeable.
1. Press any of the arrow keys to open the MAIN MENU.
2. Select Interfaces and press the right arrow key.
3. Select Serial interface and press the right arrow key.
4. Select Bit rate/Serial format/Comm. mode and press the
CHANGE key. Next specify some details (speed, format or
mode) for the options just selected. Use the up/down arrow keys
and press the SELECT key.
5. If you set RUN for communication mode, specify the interval
for RUN mode, as well. First press the SET key, adjust numbers
and change the unit with the arrow keys and finally press the
OK key.
6. Select POLL address and press SET to confirm your selection.
By using poll address, the device can be identified from other
devices connected to the same network system. Use the arrow
keys to set the poll address and finally press the OK key.
7. Press the arrow keys to select ECHO. Press ON to turn to it on.
Press OFF to turn it off.
8. Press the EXIT key to return to the basic display.
The new user port settings set using the display/keypad are effective
immediately.