Vaisala PTU300 Podręcznik Użytkownika Strona 38

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 173
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 37
User's Guide ______________________________________________________________________
36 __________________________________________________________________ M210796EN-A
Probe Mounting
In humidity measurement and especially in calibration it is essential
that temperature of the probe and measuring environment is the same.
Even a small difference in temperature between the environment and
the probe causes an error. As the curve below shows, if the
temperature is +20 °C and the relative humidity 100 %RH, a
difference of ±1 °C between the environment and the probe causes an
error of ±6 %RH.
The graph below illustrates the measurement error at 100 %RH when
the difference between the ambient and sensor temperature is 1 °C.
0507-023
Figure 19 Measurement Error at 100 %RH
Przeglądanie stron 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 ... 172 173

Komentarze do niniejszej Instrukcji

Brak uwag