User's Guide ______________________________________________________________________
138 _________________________________________________________________ M210796EN-A
Using Serial Line
Enter the ACAL command and type the multimeter reading for each
case. Continue by pressing ENTER.
ACAL
Example (current outputs):
>ACAL
Ch1 I1 (mA) ? 2.046
Ch1 I2 (mA) ? 18.087
Ch2 I1 (mA) ? 2.036
Ch2 I2 (mA) ? 18.071
>
Feeding Adjustment Information
This information is shown on the device information fields (see
sections Information Display on page 59 and Device Information on
page 87).
Using Display/Keypad
1. If you are not in the adjustment menu, press the ADJ button on
the motherboard (opens the ADJUSTMENT MENU).
2. Select Adjustment info, press the right arrow key.
3. Select Date, press the SET key. Enter date by using the arrow
keys. Press the OK key.
4. Select i, press the SET key. Enter information text including 17
characters at maximum. Use the arrow keys. Press the OK key.
5. Press the EXIT key to return to the basic display.