Chapter 3 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 37
General Instructions for Probes
with Cable
Mount the probes with a cable with the sensor head horizontally; this
way, any water condensing on the tube cannot flow onto the sensor.
0507-024
Figure 20 Horizontal Mounting of Probe
Numbers refer to Figure 20 above:
1 = To be sealed.
2 = To be insulated.
3 = Insulate the cable.
4 = Let the cable hang loosely. This prevents condensed water
running to the sensor along the cable.