When there is no alternative but to install the sensor head in the
process vertically, the point of entry must be carefully insulated. The
cable must also be allowed to hang loosely as this prevents any
condensed water from running onto the sensor head along the cable.
0507-022
Figure 21 Vertical Mounting of Probe
Numbers refer to Figure 21 above:
1 = To be sealed.
2 = Insulate the cable.
3 = To be insulated.
4 = Let the cable hang loosely. This prevents condensed water
running to the sensor along the cable.
NOTE
Please do not attach a heated probe (PTU307) to metal structures to
avoid condensation problems caused by heat conduction along the
metal.
If the process temperature is much higher than that of the
environment, the whole sensor head and preferably plenty of cable
must be inside the process. This prevents measuring inaccuracy
caused by heat conduction along the cable.
When mounted on the side of a duct or channel, the sensor head must
be inserted from the side of the duct. If this is not possible and the
sensor head must be inserted from the top, the point of entry must be
carefully insulated.
For Vaisala probe installation kits and some installation examples see
Appendix A below on page 157.