Chapter 6 ___________________________________________________Calibration and adjustment
VAISALA ______________________________________________________________________ 125
CHAPTER 6
CALIBRATION AND ADJUSTMENT
The PTU300 is fully calibrated and adjusted as shipped from factory.
Typical calibration interval is two years. Depending on the application
it may be good to make more frequent checks. Calibration must be
done always when there is a reason to believe that the device is not
within the accuracy specifications.
When defining the calibration interval the long term specifications and
the requirements of the customer must be taken into consideration.
Contact Vaisala Service Centers for details.
It is recommended that calibration and adjustment should be carried
out by Vaisala. See section Vaisala Service Centers on page 155.
Calibration and adjustment is carried out either by using the push-keys
on the motherboard, through the serial port or with the optional
display/keypad.
(Vaisala portable instruments HM70 and HMI41 can also be used).
Pressure
The user can select a simple offset or a two-point offset and gain
adjustment and use the LCI command for adjustment of pressure
transducer. The MPCI command is used for the more sophisticated
multipoint correction capability at up to eight pressure levels.
Check first what linear corrections the transmitter is currently using
before attempt to readjust the transducer. As the previous linear
corrections will disappear when new linear corrections are input, the
user has to take into account the previous linear corrections when
deciding about the new ones.