User's Guide ______________________________________________________________________
116 _________________________________________________________________ M210796EN-A
Sensor heating starts-up when the relative humidity of the measuring
environment reaches the RH-value set by a user (RH-limit). The user
can define the RH-sensor heating temperature as well as the duration
of the heating.
After the heating cycle the humidity conditions are checked and new
sensor heating is performed if the predefined conditions are reached
again.
NOTE
During the sensor heating the outputs are locked to the values
measured before the heating cycle.
Setting Humidity Sensor Heating Using
Display/Keypad
When the PTU300 leaves the factory the sensor heating follows the
factory default values. You can enable/disable the function, change the
RH-limit and define the heating temperature and duration of this
function.
1. Open the MAIN MENU by pressing any of the arrow keys.
2. Select Measuring, press the right arrow key.
3. Select Sensor heating, press the ON key.
Using Serial Line
XHEAT
Enables/disables the sensor heating.
XHEAT [xx]
where:
xx = ON / OFF
>xheat on
Extra heat : ON
>xheat off
Extra heat : OFF
>