User's Guide ______________________________________________________________________
132 _________________________________________________________________ M210796EN-A
0507-013
Figure 52 Selecting Point 1 Reference Type
6. Remove the filter from the probe and insert the probe head into
a measurement hole of the dry end reference chamber (for
example, LiCl: 11 % RH in the humidity calibrator HMK15.)
Use the adapter fitting for the PTU307 probe.
7. Wait at least 30 minutes for the sensor to stabilize. Follow the
stabilization from the GRAPH display.
8. Press the READY key when stabilized. If you have chosen the
Other reference value, enter now the reference value by using
the arrow keys.
When carrying out the 2-point adjustment proceed to the next
adjustment point and carry out the procedure as described in the
previous items.
9. Answer YES to confirm the adjustment. Press the OK key to
return to the adjustment menu.
10. Press the EXIT key to close the adjustment mode and return to
the basic display. Before closing the adjustment mode, feed the
adjustment information into the device, see section Feeding
Adjustment Information on page 138.
Using Serial Line
Note that the difference between the two humidity references must be
at least 50% RH.
1. Connect the PTU300 to a PC. See section Serial Line
Communication on page 64. Open a terminal program.
2. Carry out the chemical purge (if available).
3. Press the ADJ button.
4. Remove the filter from the probe and insert the probe head into
a measurement hole of the dry end reference chamber (for
example, LiCl: 11 % RH in the humidity calibrator HMK15).