User's Guide ______________________________________________________________________
30 __________________________________________________________________ M210796EN-A
1. Cut back outer sheath to desired length.
2. Cut back screen braiding or screen foil to dimension X (see
figure 3).
3. Push the domed cap nut (item 1) and the seal insert with contact
socket of the gland (item 2+3) onto the cable as shown in the
diagram.
4. Bend over the screen braiding or screen foil by about 90º (item
4).
5. Push the seal insert with the contact socket of the gland (item
2+3) up to the screen braiding or screen foil.
6. Mount lower part (item 5) on the housing.
7. Push the seal with the contact socket of the gland and (item 2+3)
flush into the lower part (item 5).
8. Screw the domed cap nut (item 1) onto the lower part (item 5).
Grounding the Transmitter Housing
In case you need to ground the transmitter housing, the grounding
connector is found inside the housing, see Figure 2 on page 17 above.
Note anyhow that the probe head is connected to the same potential as
the housing. Make sure that different groundings are made to the same
potential. Otherwise harmful ground currents may be generated.
If it is needed to have galvanic isolation of the power supply line from
the output signals, PTU300 can be ordered with optional output
isolation module. This module prevents harmful grounding loops.