Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 71
Getting Measurement Message
from Serial Line
Enter the R command to start output of measurements. Enter the S
command, press the Esc key or reset the transmitter to stop outputting.
See command SMODE to change the default (power-up) operation
mode.
Format the output by using the following commands:
- outputting interval can be changed with the INTV command.
- output message format can be changed with a command FORM.
Example:
>r
P= 1021.6 hPa T= 23.2 'C RH= 5.8 %RH
>
Stopping Continuous Outputting
S
Use the S command to end the RUN mode. After this command all
other commands can be used.
Outputting Reading Once
SEND
Use the SEND command to output the reading once in STOP mode:
The output format depends on which parameters the transmitter can
output.
Examples:
P= 1021.6 hPa T= 23.3 'C RH= 5.7 %RH
>
If value is too long to fit to the allocated space or if there is an error in
outputting the quantity, value is displayed with stars ‘*’.