User's Guide ______________________________________________________________________
128 _________________________________________________________________ M210796EN-A
5. Let the readings stabilize. Press the READY key when
stabilized.
6. Enter the actual pressure of the reference used using the
up/down arrow keys. Press the OK key.
7. Press the YES key to perform the adjustment. Press the OK to
return to the adjustment menu.
1-point Adjustment Using Serial
Line
NOTE
Making adjustments is possible only after adjustments are unlocked.
To unlock the adjustment menu, press the ADJ button on the
motherboard of the transmitter.
LCI
Use the command LCI
- to activate or deactivate the linear adjustment function
- to enter new linear offset and offset/gain pressure corrections to
the transmitter
- to edit existing linear offset and offset/gain pressure corrections.
Key in the linear corrections for each pressure transducer separately.
NOTE
The new linear corrections will always cancel the previous
corrections as well as the valid date of calibration of the transmitter.
Example:
>lci
P1 linear adj. : OFF ? on
P1 1.reading : 0.000 ? 950
P1 1.correction: 0.000 ? 0.22
P1 2.reading : 0.000 ? 1120
P1 2.correction: 0.000 ? 0.15