User's Guide ______________________________________________________________________
50 __________________________________________________________________ M210796EN-A
Table 7 Connecting the Twisted Pair Wires to the Screw
Terminals
Screw terminal Data line
(2-wire RS-485)
Data line
(4-wire RS-485/422)
1 (not connected) RxB
2 (not connected) RxA
3 Data pair shield Data pair shield
4 B TxB
5 A TxA
6. If you use RS-485 (or RS-422) to connect just one PTU300 to a
master computer, enable the internal termination of PTU300 by
switching switches 1 and 2 ON. Make sure that the master's end
of the line is also terminated (by using master's internal
termination or with a separate terminator).
If you are connecting many transmitters to the same RS-485 bus,
make sure that switches 1 and 2 are OFF and terminate the bus
with separate terminators at both ends. This allows removing
any transmitter without blocking the bus operation.
NOTE
If you use the internal termination of the transmitter at the end of the
RS-485 bus (instead of using separate terminators) removing that
transmitter will block the bus operation.
7. Use the bus type (4-wire/2-wire) to select the selection switch 3.
In 4-wire mode RS-485 master sends data to the PTU300
through terminals RxA and RxB and receives data from PTU300
through terminals TxA and TxB.