Vaisala PTU300 Podręcznik Użytkownika Strona 94

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 186
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 93
USER'S GUIDE____________________________________________________________________
92 __________________________________________________________________ M210796EN-E
Formatting Serial Line Message
FTIME and FDATE
FTIME and FDATE commands will enable/disable output of time and
date to the serial line. To add time to R and SEND outputs enter:
FTIME [x]
To add date to R and SEND outputs enter:
FDATE [x]
where
x = ON or OFF
Example:
>send
RH= 98.4 %RH T= 31.0 'C
>ftime on
Form. time : ON
>send
03:47:59 RH= 98.4 %RH T= 31.0 'C
>fdate on
Form. date : ON
>send
2004-07-05 03:48:03 RH= 98.4 %RH T= 31.0 'C
>
General Settings
Changing Quantities and Units
Use serial commands or the optional display/keypad to change quantities
and units. For more information on available quantities and units, see
Table 4 on page 19. For more information on optional quantities, see
Table 5 on page 19.
NOTE
Only the quantities selected when ordering the device can be selected as
a display output quantity.