DSEL [xxx]
where
xxx = Data recording quantity. For more information on available
quantities and units, see Table 4 on page 19. For more
information on optional quantities, see Table 5 on page 19.
Example:
>dsel rh t tdf
RH T Tdf
>
Enter the command without parameters and press ENTER to display
current recording parameters.
View Recorded Data
If the device is provided with the optional display, the graphical display
shows the data of the selected quantities, one at a time. For details about
graphical display, see section Graphic History on page 65.
You may also dump the logged data to the serial line in numeric form
with the following commands.
DIR
Use the serial line and enter the DIR command to check the available
files.
Without the data logger module, the device records six files (six
observation periods) for each selected quantity. The data logger raises the
number of recorded files to seven for each quantity. Thus, the total
amount of the files varies between 6 and 21. See Table 13 on page 65.
Select, fo
r example, two quantities (P and T). The last column illustrates
the number of data points that has been stored in the file.
Example (data logger module installed):