If a value is too long to fit to the allocated space in the output, or if there
is an error in outputting the quantity, the value is displayed with stars '*'.
Example:
RH=***.* %RH T= 31.0 'C
You can change the format of the output with the following commands:
- Outputting interval can be changed with the INTV command.
- Output message format can be changed with the FORM command.
- Status of chemical purge and probe heating can be added with the
FST command.
- Date and time information can be added with commands FDATE and
FTIME
Stopping Continuous Outputting
S
Use the S command to end the RUN mode. After this command all other
commands can be used. You can also press the Esc button or reset the
transmitter to stop the outputting.
See command SMODE to change the default (power-up) operation
mode.
Outputting Reading Once
SEND
Use the SEND command to output the reading once in STOP mode:
The output format depends on which parameters the transmitter can
output.
Example:
>send
P= 1021.6 hPa T= 23.3 'C RH= 5.7 %RH
>