USER'S GUIDE____________________________________________________________________
64 __________________________________________________________________ M210796EN-E
where:
6 = Decreasing, then steady; or decreasing, then decreasing more
slowly; atmospheric pressure now lower than three hours ago
7 = Decreasing (steadily or unsteadily); atmospheric pressure now
lower than three hours ago
8 = Steady or increasing, then decreasing; or decreasing then
decreasing more rapidly; atmospheric pressure now lower than
three hours ago
Source: The World Meteorological Organization (WMO) publication
Manual on Codes Vol. I.1, International Codes, Part A - Alphanumerical
Codes, 1995 Edition, WMO - No. 306. Section C, Code Table 0200: a.
Using Serial Line
Pressure 3h trend and tendency reading is also available through serial
line. Key in the lines below:
>form "P=" p "trend=" p3h "tend=" a3h #r#n
OK
>send <cr>
P= 1024.7trend= 0.8tend=1
The last line shows the values.
For more information on the Form command, see the section Changing
Quantities and Units, starting on page 92.
Missing trend
In addition to this the PTU300 series barometers output a code "*" when
the pressure tendency has not yet been calculated that is, less than three
hours have elapsed since the power-up of the barometer. The absence of
the pressure trend is indicated in a similar manner, too.
NOTE
When the P
3H
tendency is chosen as a display quantity, the PTU300
logs actual measured pressure, instead of trend or tendency value.