Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 87
If PuTTY is unable to connect the IP address you entered, it will
show you an error message instead. If this happens, check the
IP address and the connections, restart PuTTY, and try again.
List of Serial Commands
The bold text in the brackets is the default setting. To issue a command,
type it on your computer and press the Enter key.
Table 19 Measurement Commands
Command Description
R Start the continuous outputting
S Stop the continuous outputting
INTV [0 ... 255 S/MIN/H] Set the continuous output interval
(for RUN mode)
SEND [0 ... 99] Output the reading once
SEND D Outputting the reading with the raw data
SCOM Define a user-specific SEND command for one
message output
SMODE
[STOP/RUN/POLL]
Set the serial interface mode
SDELAY View or set user port (RS232 or RS485)
answer minimum delay
SERI [baud p d s] User Port settings (Default: 4800 E 7 1) baud:
300 ... 115200
ADDR [0 ... 99] Set the transmitter address (for POLL mode)
NET View or set networking parameters for LAN and
WLAN interfaces
WLAN View or set wireless network parameters for
WLAN interface
OPEN [0 ... 99] Open a temporary connection to a POLL mode
device
CLOSE Close the temporary connection (Back to POLL
mode)
Table 20 Formatting Commands
Command Description
FORM Set the output format of SEND and R
commands
TIME Set the time
DATE Set the date
UNIT Select the metric or non-metric output units
Table 21 Data Recording Commands
Command Description
DIR Display recorded files
PLAY [0 ... 21] [START END] Output recorded data file. Start and end times
can only be specified if the data logger module
is installed. The times must be given in the
following format:
yyyy-mm-dd hh:mm:ss