Vaisala PTU300 Podręcznik Użytkownika Strona 154

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 186
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 153
USER'S GUIDE____________________________________________________________________
152 _________________________________________________________________ M210796EN-E
Using Display/Keypad
1. Press the ADJ button to open the ADJUSTMENT MENU.
2. Select Adjust analog outputs, press the right arrow button.
3. Select the output to be adjusted Adjust analog output 1/2, press
the START key.
4. Measure the first analog output value with a multimeter. Give the
measured value by using the arrow buttons. Press the OK key.
5. Measure the second analog output value with a multimeter. Give
the measured value by using the arrow buttons. Press the OK key.
6. Press the OK key to return to the adjustment menu.
7. Press the EXIT key to close the adjustment mode and to return to
the basic display.
Using Serial Line
Enter the ACAL command and type the multimeter reading for each
case. Continue by pressing ENTER.
ACAL
Example (current outputs):
>ACAL
Ch1 I1 (mA) ? 2.046
Ch1 I2 (mA) ? 18.087
Ch2 I1 (mA) ? 2.036
Ch2 I2 (mA) ? 18.071
>
Przeglądanie stron 153
1 2 ... 149 150 151 152 153 154 155 156 157 158 159 ... 185 186

Komentarze do niniejszej Instrukcji

Brak uwag