Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 125
RTEST
Use the serial line command RTEST to test the operation of the relays.
RTEST [x1 x2 x3 x4]
where
x
= ON/OFF
Example: Activate and then release all four relays.
>rtest on on on on
ON ON ON ON
>
>rtest off off off off
OFF OFF OFF OFF
>
Enter the command RTEST without parameters to stop testing.
Operation of the RS-485 Module
RS-485 interface enables communication between RS-485 network and
PTU300 transmitter. The RS-485 interface is isolated and offers a
maximum communications rate of 115 200 bits/s. (For maximum bus
length of 1 km, use bit rate 19200 b/s or less.)
When selecting an RS-232-RS-485 converters for the network, avoid self
powered converters as they don't necessarily support the needed power
consumption.
Echo function shall be always disabled (OFF) when using the 2-wire
connection. When using the 4-wire connection you can disable/enable the
echo setting.
NOTE
User port on PTU300 main board cannot be used and connected when
RS-485 module is connected. Service port is operating normally.