12 13
°C
°C
0
0-80 -60 -40 -20 20 40 60 80 100 120 140 160
180
-0.7
-0.6
-0.5
-0.4
-0.3
-0.2
-0.1
0.1
0.2
0.3
0.4
0.5
0.6
0.7
Temperature sensor PT 100 RTD 1/3 Class B IEC 751
Other variables available (depends on model)
dewpoint temperature, mixing ratio, absolute humidity, wet bulb
temperature, enthalpy, water vapor pressure
Performance
RELATIVE HUMIDITY
Measurement range 0 ... 100 %RH
Accuracy (including non-linearity, hysteresis, and repeatability)
with Vaisala HUMICAP
®
180 or 180R for typical applications
with Vaisala HUMICAP
®
180C or 180RC for applications with
chemical purge/warmed probe
at +15 ... +25 °C (59 ... +77 °F) ±1 %RH (0 ... 90 %)
±1.7 %RH (90 ... 100 %RH)
at -20 ... +40 °C (-4 ... +104 °F) ±(1.0 + 0.008 x reading) %RH
at -40 ... +180 °C (-40 ... +356 °F) ±(1.5 + 0.015 x reading) %RH
Factory calibration uncertainty (+20 °C) ± 0.6 %RH (0 ... 40 %RH)
± 1.0 %RH (40 ... 97 %RH)
(Defined as ±2 standard deviation limits. Small
variations possible, see also calibration certificate.)
Response time (90 %) at +20 °C (+68 °F) 8 s / 17 s* with grid filter
in still air 20 s / 50 s* with grid + steel netting filter
40 s / 60 s* with sintered filter
* with HUMICAP
®
180R or 180RC sensor
TEMPERATURE
Accuracy at +20 °C (+68 °F) ± 0.2 °C (± 0.36 °F)
Accuracy over temperature range (measurement range
depends on model)
Inputs and outputs
Operating voltage 10 ... 35 VDC, 24 VAC
with optional power supply module 100 ... 240 VAC, 50/60 HZ
Power consumption at +20 °C (U
in
24 VDC)
RS-232 max. 25 mA
U
out
2 x 0 ... 1 V/0 ... 5 V/0 ... 10 V max. 25 mA
I
out
2 x 0 ... 20 mA max. 60 mA
display and backlight + 20 mA
during chemical purge max. 110 mA
during probe heating (HMT337) + 120 mA
Analog outputs (2 standard, 3rd optional)
current output 0 ... 20 mA, 4 ... 20 mA
voltage output 0... 1 V, 0... 5 V, 0... 10 V
Accuracy of analog outputs at +20 °C ±0.05% full scale
Temperature dependence of the
analog outputs ±0.005%/°C full scale
External loads
current outputs R
L
< 500 ohm
0 ... 1 V output R
L
> 2 kohm
0 ... 5 V and 0 ... 10 V outputs R
L
> 10 kohm
Max. wire size 0.5 mm
2
(AWG 20)
stranded wires recommended
Digital outputs RS-232, RS-485 (optional)
Service connection RS-232, USB
Relay outputs (optional) 0.5 A, 250 VAC
Technical data
Ethernet interface (optional)
Supported standards 10/100Base-T
Connector RJ45
Protocols Telnet
Software support Vaisala MI70 link
WLAN interface (optional)
Supported standards 802.11b
Antenna connector type RP-SMA
Protocols Telnet
Security WEP 64/128,WPA
Software support Vaisala MI70 link
Authentication / Encryption (WLAN)
Open / no encryption
Open / WEP
WPA Pre shared key / TKIP
WPA Pre shared key / CCMP (a.k.a. WPA2)
Optional data logger with real-time clock
Logged parameters max. three with trend/min/max values
Logging interval 10 sec (fixed)
Max. logging period 4 years 5 months
Logged points 13.7 million points per parameter
Battery lifetime min. 5 years
Display LCD with backlight, graphic trend display
of any parameter
Menu languages English, Chinese, Japanese, Spanish, German,
French, Swedish, Russian, Finnish
Mechanics
Cable bushing M20 x 1.5 for cable diameter
Measurement range 8 ... 11 mm/0.31 ... 0.43"
Conduit fitting 1/2" NPT
User cable connector (optional) M12 series 8-pin (male)
option 1 female plug with 5 m (16.4 ft) black cable
option 2 female plug with screw terminals
Probe cable diameter
HMT333 (+80 °C) 6.0 mm
other probes 5.5 mm
Housing material G-AlSi 10 Mg (DIN1725)
Housing classification IP 65 (NEMA4)
Operating environment
Operating temperature
for probe same as measurement range
for transmitter body -40 ... +60 °C (-40 ... 140 °F)
with display 0 ... +60 °C (32 ... 140 °F)
Electromagnetic compatibility Complies with EMC standard
EN61326-1, Industrial Environment
Dimensions
Dimensions in mm (inches)
Mounting options
Mounting with Wall
Mounting Kit
Pole Installation with Installation
Kit for Pole or Pipeline
Mounting with DIN Rail
Installation Kit
Mounting Rain Shield with
Installation Kit
106 (4.17)
116 (4.57)
183 (7.20)
Transmitter with WLAN antenna
TYPE APPROVED PRODUCT
CERTIFICATE NO.: A-11440
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Ref. B210951EN-A ©Vaisala 2010
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
Komentarze do niniejszej Instrukcji