www.vaisala.com
96 97
MM70 Hand-Held Moisture and Temperature
Meter for Spot-Checking in Oil
The Vaisala HUMICAP
®
Hand-held
Moisture Meter for Oil MM70 enables
reliable detection of moisture in oil.
In-line process checking
through ball valve
The probe can be inserted directly
into the process pipe through a ball
valve without draining the oil in the
system.
Water activity
measurement
The MM70 measures moisture in oil
in terms of the water activity (aw)
and temperature (T). Water activity
directly indicates whether there is
a risk of free water formation. The
measurement is independent of oil
type, age and temperature.
ppm calculation included
The MM70 has an embedded model
for expressing moisture as ppm in
mineral transformer oil. The customer
can enter up to three other oil models
into the meter’s memory.
The MM70 is an ideal
tool for the preventive
maintenance of oil-filled
systems. The water
activity measurement
indicates the margin to
free water formation,
which causes severe
problems in lubrication
systems.
Numerical and
graphical display
The MM70 features a multilingual,
menu-based user interface and a
backlit LCD display. The measurement
parameters can be numerically and
graphically displayed and logged
into the meter's memory at the same
time. An analog output option is also
available.
Connection to PC
The optional MI70 Link Windows
®
software in combination with a
USB connection cable is used to
transfer logged data and real time
measurement data from the MM70
to a PC.
Proven Vaisala HUMICAP®
technology
The MM70 incorporates the latest
generation of the Vaisala HUMICAP
®
Sensor, developed for demanding
moisture measurements in liquid
hydrocarbons. The sensor’s excellent
chemical tolerance provides accurate
and reliable measurement over the
measurement range.
Speedy service -
once a year
The meter can be recalibrated
by sending the probe to Vaisala
Service, or customers can calibrate
the instrument themselves using a
standard relative humidity calibration.
Multi-probe operation
One or two probes can be connected
simultaneously. Maintenance teams
can use additional Vaisala dewpoint
or relative humidity probes for other
tasks. For example, a dewpoint probe
is ideal for checking the moisture
inside washed and dried oil tanks.
Features/Benefits
▪
Measurement independent of
oil type, age and temperature
▪
In-line process checking
through ball valve, no need to
drain the oil
▪
Rugged and reliable
construction
▪
Excellent pressure and
temperature tolerance
▪
Data can be logged and
transferred to a PC
▪
Vaisala HUMICAP® Sensor -
proven field performance
since 1973
▪
Compatible with Vaisala’s fixed
oil moisture instruments
▪
No reference oil needed for
recalibration
▪
NIST traceable (certificate
included)
Technical data
Performance
WATER ACTIVITY
Measurement range a
w
0 ... 1
Accuracy (including nonlinearity, hysteresis and repeatability)
When calibrated against salt solutions (ASTM E104-85):
0 ... 0.9 ±0.02
0.9 ... 1.0 ±0.03
Maximum achievable accuracy when calibrated against
high-quality, certified humidity standards:
0 ... 0.9 ±0.01
0.9 ... 1.0 ±0.02
Response time (90%) at +20 °C (+68 °F)
in still oil (with stainless steel filter) 10 min.
Sensor Vaisala HUMICAP
®
Recommended recalibration interval 1 year
TEMPERATURE
Measurement range -40 ... +100 °C (-40 ... +212 °F)
Typical accuracy at +20 °C ±0.2 °C (±0.36 °F)
Typical temperature dependence
of electronics ±0.005 °C/°C (±0.005 °F/°F)
Sensor Pt 100 IEC 751 1/3 class B
Typical long-term stability better than 0.01 aw / year
Options and accessories
Carrying case MI70CASE2
Ball valve set (incl. fitting body & blanking plug) HMP228BVS
Probe cable extension, 10 m 213107SP
Transmitter connection cables for
HMP228 27159ZZ
MMT318 DRW216050
MMT/HMT330/DMT340 211339
MI70 Link software with USB cable 219687
MI70 Link software with serial port cable MI70LINK
Analog output cable 27168ZZ
Sensor protection HM47453SP
Dewpoint measurement probes DMP74A/B
Relative humidity measurement probes HMP75, HMP76, HMP77
Dimensions
Dimensions in mm (inches)
Inputs and outputs
Power supply Rechargeable NiMH battery pack with
AC-adapter or 4xAA-size alkalines, type IEC LR6
Battery operation time
continuous use 48 h typical at +20 °C (+68 °F)
data logging use up to a month, depending on
logging interval
Menu languages English, Chinese, Spanish, French, German,
Japanese, Russian, Swedish, Finnish
Operating environment
PROBE
Operating temperature range for electronics -40 ... +60 °C
(-40 ... +140 °F)
Operating pressure range max. 20 bar
during installation through ball valve max. 10 bar
Oil flow range max. 1 m/s
INDICATOR
Operating temperature range -10 ... +40 °C (+14 ... +104 °F)
Operating humidity range non-condensing
ELECTROMAGNETIC COMPATIBILITY
Complies with EMC standard EN61326-1, Electrical equipment for
measurement, control and laboratory use - EMC requirements;
Portable equipment.
Ref. B210960EN-A ©Vaisala 2010
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Mechanics
PROBE
Housing classification IP65 (NEMA 4)
Housing material ABS/PC blend
Probe material Stainless steel (AISI316L)
Cable length between probe and indicator 1.9 m,
10 m extension available
Weight 506 g
INDICATOR
Housing classification IP54
Weight 400 g
Probe inputs 1 or 2
Display LCD with backlight, graphic trend display of
any parameter, character height up to 16 mm
Analog output 0 ... 1 VDC
Output resolution 0.6 mV
PC interface MI70 Link software with USB or serial port cable
Data logging capacity 2700 points
Alarm Audible alarm function
HUMICAP
®
is a registered trademark of Vaisala.
R
1
/
2
" ISO 7/1
Adjustment range 250 mm