www.vaisala.com
134 135
Vaisala Barometric Pressure Transfer
Standard PTB330TS
Performance
Barometric pressure (PTB330)
Measurement range 500 ... 1100 hPa
Linearity* ±0.05 hPa
Hysteresis* ±0.03 hPa
Repeatability* ±0.03 hPa
Calibration uncertainty** ±0.07 hPa
Accuracy at +20 °C (+68 °F) *** ±0.10 hPa
Temperature dependence**** ±0.1 hPa
Total accuracy -40... +60 °C (-40...+140 °F) ±0.15 hPa
Long-term stability ±0.1 hPa/year
Settling time at power-up (one sensor) 4 s
Response time (one sensor) 2 s
Acceleration sensitivity negligible
* Defined as ±2 standard deviation limits of endpoint nonlinearity, hysteresis or
repeatability error.
㻖㻖㻌 㻰㼑㼒㼕㼚㼑㼐㻌㼍㼟㻌㼼㻞㻌㼟㼠㼍㼚㼐㼍㼞㼐㻌㼐㼑㼢㼕㼍㼠㼕㼛㼚㻌㼘㼕㼙㼕㼠㼟㻌㼛㼒㻌㼕㼚㼍㼏㼏㼡㼞㼍㼏㼥㻌㼛㼒㻌㼠㼔㼑㻌㼣㼛㼞㼗㼕㼚㼓㻌㼟㼠㼍㼚㼐㼍㼞㼐㻌㼕㼚㼏㼘㼡㼐㼕㼚㼓㻌
㼠㼞㼍㼏㼑㼍㼎㼕㼘㼕㼠㼥㻌㼠㼛㻌㻺㻵㻿㼀㻚
㻖㻖㻖㻌 㻰㼑㼒㼕㼚㼑㼐㻌㼍㼟㻌㼠㼔㼑㻌㼞㼛㼛㼠㻌㼟㼡㼙㻌㼛㼒㻌㼠㼔㼑㻌㼟㼝㼡㼍㼞㼑㼟㻌㻔㻾㻿㻿㻕㻌㼛㼒㻌㼑㼚㼐㼜㼛㼕㼚㼠㻌㼚㼛㼚㻙㼘㼕㼚㼑㼍㼞㼕㼠㼥㻘㻌㻌㼔㼥㼟㼠㼑㼞㼑㼟㼕㼟㻌㼑㼞㼞㼛㼞㻘㻌
㼞㼑㼜㼑㼍㼠㼍㼎㼕㼘㼕㼠㼥㻌㼑㼞㼞㼛㼞㻌㼍㼚㼐㻌㼏㼍㼘㼕㼎㼞㼍㼠㼕㼛㼚㻌㼡㼚㼏㼑㼞㼠㼍㼕㼚㼠㼥㻌㼍㼠㻌㼞㼛㼛㼙㻌㼠㼑㼙㼜㼑㼞㼍㼠㼡㼞㼑㻚
㻖㻖㻖㻖㻌㻰㼑㼒㼕㼚㼑㼐㻌㼍㼟㻌㼼㻞㻌㼟㼠㼍㼚㼐㼍㼞㼐㻌㼐㼑㼢㼕㼍㼠㼕㼛㼚㻌㼘㼕㼙㼕㼠㼟㻌㼛㼒㻌㼠㼑㼙㼜㼑㼞㼍㼠㼡㼞㼑㻌㼐㼑㼜㼑㼚㼐㼑㼚㼏㼑㻌㼛㼢㼑㼞㻌㼠㼔㼑㻌㼛㼜㼑㼞㼍㼠㼕㼚㼓㻌
㼠㼑㼙㼜㼑㼞㼍㼠㼡㼞㼑㻌㼞㼍㼚㼓㼑㻚
Relative humidity (HMP155)
Measurement range 0 ... 100 %RH
Accuracy (incl. non-linearity, hysteresis and
repeatability) at +15 ... +25 °C (+59 ... +77 °F)
±1 %RH (0 ... 90 %RH)
±1.7 %RH (90 ... 100 %RH)
-10 ... +40 °C (-4 ... 104 °F) ±(1.0 + 0.008 x reading) %RH
Factory calibration uncertainty (+20 °C /+68 °F)
±0.6 %RH (0 ... 40 %RH)*
±1.0 %RH (40 ... 97 %RH)*
Humidity sensor HUMICAP180R
HUMICAP180RC
Response time at +20 °C in still air with a sintered PTFE filter
63 % 20 s
90 % 60 s
* Defined as ±2 standard deviation limits. Small variations possible, see also calibration
certificate.
Technical data
These specications apply when MI70, PTB330 and
HMP155 are used together in the PTB330TS product.
For individual specications, please refer to the product
documentation and brochures of the PTB330 and HMP155.
General
Operating temperature range -10 ... +40 °C (+14 ... +104 )°F
Operating humidity range non-condensing
Maximum pressure limit 5000 hPa abs.
Power supply Rechargeable NiMH battery pack
with AC-adapter or 4xAA-size
alkalines, type IEC LR6
Operation time
(using rechargeable battery pack)
Continuous use with PTB330 11 h typical at +20 °C (+68 °F)
Datalogging use up to 30 days
Menu languages English, Chinese, French, Spanish,
German, Russian, Japanese,
Swedish, Finnish
Display LCD with backlight, graphic
trend display of any parameter,
character height up to 16 mm
Data logging capacity 2700 points
Alarm audible alarm function
PTB330TS is in conformity with the following EU directives:
- EMC Directive (2004/108/EC) Complies with the EMC product family standard EN61326-1,
Electrical equipment for measurement control and laboratory use - Basic immunity test
requirements.
- Low Voltage Directive (2006/95/EC)
- ROHS Directive (2002/95/EC)
Temperature (HMP155)
Measurement range -10 ... +40 °C (+14 ... +104 °F)
Accuracy
-10 ... +20 °C ±(0.176 - 0.0028 x
temperature) °C
+20 ... +40 °C ±(0.07 + 0.0025 x
temperature) °C
Accuracy over temperature range (see graph overleaf)
Temperature sensor Pt100 RTD 1/3 Class B IEC 751
Response time with additional temperature probe in 3 m/s air flow
63 % <20 s
90 % <35 s
Technical data
Inputs and outputs
MI70 probe ports 2
MI70 data interface RS-232 (accessible only with MI70
Link software)
PTB330 supply voltage 10 ... 35 VDC (if not powered by
MI70)
PTB330 data interface RS-232C
PTB330 serial I/O connectors RJ45 (service port)
Male 8-pin M12 (user port)
HMP155 data interface RS-485
HMP155 serial I/O connector Male 8-pin M12
Mechanics
PTB330
Housing material G-AlSi 10 Mg (DIN 1725)
Housing classification IP65
Pressure connector M5 (10-32) internal thread
Pressure fitting barbed fitting for 1/8” I.D.
tubing or quick connector with
shutoff valve for 1/8” hose
HMP155
Housing material PC
Housing classification IP66
Additional T-probe cable length 2 m
Cable material PUR
Sensor protection Sintered PTFE
MI70 MEASUREMENT INDICATOR
Housing classification IP54
Housing material ABS/PC blend
TRANSPORT CASE
Housing classification
(when closed) IP67
Plastic parts TTX01
®
, PP+SEBS, POM
Metal parts stainless steel AISI303
Interior foam material PE and polyether
Weight with all instruments and
typical accessories 5.9 kg
Exterior dimensions (LxWxH) 405×330×165 mm
(15.94×12.99×6.50) inch
Accessories
PTB330
MI70 – PTB330 Spiral Cable 223235SP
USB-RJ45 serial connection cable 219685
Serial connection cable 19446ZZ
Barbed fitting 1/8” 19498SP
Quick Connector 1/8” 220186
Transport case with interior foams
and tabletop casing for PTB330 224068SP
Tabletop casing for PTB330 224064SP
MI70
USB cable for MI70, 219687
includes MI70 Link software
MI70 Link software MI70LINK
MI70 connection cable to HMT330,
MMT330, DMT340, HMT100, PTB330 211339
MI70 battery pack 26755
variety of AC adapters available
HMP155
HMP155 – MI70 connection cable 221801
Protection set for HMP155 calibration
buttons: protective cover,
2 O-rings and protective plug 221318
USB cable for HMP155 221040
Sintered teflon filter + O-ring 219452SP
Humidity sensor HUMICAP180R
Humidity Calibrator HMK15
Available Parameters
Pressure parameters P, P3h, HCP, QFE, QNH
Humidity and temperature RH, T, Tdf, Td, x, Tw
parameters
Ref. B210786EN-A ©Vaisala 2009
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Accuracy of HMP155 temperature measurement over
temperature range
Komentarze do niniejszej Instrukcji