www.vaisala.com
42 43
12
(0.47)
71
(2.8)
M
12
x1
Vaisala INTERCAP
Humidity and Temperature
Probe HMP60
HMP60
The HMP60 is a simple, durable and
cost-effective humidity probe. It is
suitable for volume applications,
integration into other manufacturers
equipment, incubators, glove boxes,
greenhouses, fermentation chambers,
and data loggers.
Easy installation
The probe cable has a screw-on
quick connector for easy installation.
Different cable lengths are available.
Also other compatible M8 series cables
can be used. Accessories are available
for different installation needs.
Low current consumption
The HMP60 is suitable for battery-
powered applications because of its
very low current consumption.
The HMP60 for extreme conditions.
Several outputs
There are two congurable voltage
outputs with relative humidity,
temperature or dew point scaling.
Four voltage output ranges are
available.
Rugged design
The HMP60 is designed for extreme
conditions. The aluminum body of
the HMP60 is classied as IP65.
The probe has a sealed structure
and the sensor is protected by a
membrane lter and a plastic grid,
or optionally by a stainless steel lter.
Recalibration not needed
The Vaisala INTERCAP
®
Sensor is
interchangeable. No recalibration is
required; the sensor can simply be
replaced, also in the eld.
Features/Benefits
Miniature-size humidity probe
Low power consumption
Measurement range:
0 ... 100 %RH; -40 ... +60°C
Cable detachable with
standard M8 connector
Rugged metal housing
Interchangeable Vaisala
INTERCAP
Sensor
Optional RS485 digital output
Optional dew point output
Applications: volume
applications, integration
into other manufacturers’
equipment, glove boxes,
greenhouses, fermentation
chambers, data loggers
Ref. B210851EN-A ©Vaisala 2010
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Performance
RELATIVE HUMIDITY
Measurement range 0 ... 100 %RH
Typical accuracy
temperature range 0 ... +40 °C
0 ... 90 %RH ±3 %RH
90 ... 100 %RH ±5 %RH
temperature range -40 ... 0 °C, +40 ... +60 °C
0 ... 90 %RH ±5 %RH
90 ... 100 %RH ±7 %RH
Humidity sensor Vaisala INTERCAP
®
TEMPERATURE
Measurement range -40 ... +60 °C
Accuracy over temperature range -40 ... +60 °C ±0.6 °C
DEW POINT
Measurement range -40 ... +60 °C
Typical accuracy
temperature range 0 ... +40 °C
when dew point depression < 15 °C ±2 °C
temperature range -40 ... 0 °C, +40 ... +60 °C
when dew point depression < 10 °C ±3 °C
dew point depression = ambient temperature - dew point
Inputs and outputs
Operating voltage
(Use lowest available operating
voltage to minimize heating.)
5 ... 28 VDC / 8 ... 28 VDC with
5 V output
8 ... 28VDC with loop power
converter
Current consumption 1 mA average, max. peak 5 mA
Start-up time at operating voltage <14 V: 1 s; >14 V: 4 s
Outputs
2 channels 0 ... 1 VDC/0 ... 2.5 VDC / 0 ... 5 VDC/1 ... 5 VDC
1-channel loop-power converter (separate
module, compatible with humidity accuracy only) 4 ... 20 mA
digital output (optional) RS485 2-wire half duplex
Mechanics
Materials
body chrome coated aluminum
grid filter chrome coated ABS plastic
cable polyurethane or FEP
Housing classification IP65
Body thread MI2x1 / 10 mm
Cable connector 4-pin M8 (IEC 60947-5-2)
Weight with 0.3 m cable 23 g
Operating environment
Operating temperature -40 ... +60 °C
Electromagnetic compatibility Complies with EMC standard
EN61326-1, Basic immunity test
requirements
Options and accessories
Vaisala INTERCAP
®
Sensor, 1 piece
Vaisala INTERCAP
®
Sensor, 10 pcs
15778HM
INTERCAPSET-10PCS
Sensor protection
plastic grid DRW010522
membrane filter DRW010525
stainless steel sintered filter HM46670SP
4 ... 20mA loop power converter UI-CONVERTER-1
Mounting bracket with lid 225979
Plastic M12 installation nuts, pair 18350SP
USB cable for PC connection 219690
Probe mounting clamp set, 10 pcs 226067
Probe mounting flange 226061
Connection cables
0.3 m PU HMP50Z032
3 m PU HMP50Z30A
180°C 3 m FEP 226902
Dimensions
in mm (inches)
External loads
0 ... 1 V R
L
min 10 k
0 ... 2.5 V /0 ... 5 V R
L
min 50 k
Technical data