Chapter 3 _________________________________________________________ Selecting Location
VAISALA_______________________________________________________________________ 49
CHAPTER 3
SELECTING LOCATION
This chapter provides information on siting the station and the sensors
correctly.
Siting the Station
Finding a suitable site for the weather station is important for getting
representative ambient measurements. Normally, the suitable site
should represent the general area of interest. When locating the weather
station, consider the items presented in the sections for each sensor. The
descriptions are not exhaustive, for further information refer to local
and WMO recommendations.
Combined Wind Sensor
Allow sufficient clearance for the wind sensor. Install the wind sensor
away from buildings or any other objects that might affect the airflow.
In general, any object of height (h) will not remarkably disturb the wind
measurement at a minimum distance of 10 × h. There should be at least
150 m (500 ft.) open area to all directions from the mast. Refer to Figure
35 on page 50.