Chapter 5 ______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 53
Changing the HUMICAP 180R Sensor
(HMP110)
NOTE
Replacing the humidity sensor of the HMP110 is not necessary in
normal operation. If the accuracy of the probe does not seem to be
within specification, it is likely that the accuracy can be restored by
performing the adjustment procedure. See section Two-point
Adjustment of RH Measurement (HMP110) on page 48.
Follow this procedure to replace the humidity sensor of the HMP110 in
case it has been damaged, or normal adjustment is not sufficient to
restore the measurement accuracy. Calibration and adjustment of the
humidity measurement is required after the sensor change.
To perform this procedure, you need a new HUMICAP 180R sensor. It is
also recommended that you replace the filter with a new one. For order
codes, see section Options and Accessories on page 60.
1. Remove the filter from the probe by turning it counter-clockwise.
2. Pull out the old HUMICAP 180R sensor (the larger of the two
sensors inside the probe) and insert a new one. Refer to Figure 14
on page 52. Handle the new sensor by the plastic socket. DO NOT
TOUCH THE SENSOR PLATE.
3. Perform a two-point adjustment of the RH measurement as
instructed in section Two-point Adjustment of RH Measurement
(HMP110) on page 48.
4. Attach a new filter on the probe.