Changing the Filter
The filter on the probe should be replaced when it is damaged or dirty.
1. Turn the filter counter-clockwise to loosen it.
2. Remove the filter from the probe. Be careful not to touch the
sensors with the filter. Without the filter in place, the sensors are
easily damaged – handle the probe carefully.
3. Install a new filter on the probe, and tighten it so it is finger-tight.
Make sure the filter sits straight and meets the threads properly.
New filters can be ordered from Vaisala. For available filters, see section
Filter Options on page 15. The order codes of the filters are listed in
section Options and Accessories on page 60.
Calibration Procedure
To calibrate your probe, you need a known stable humidity or
temperature reference, and a way to read the output of the probe (analog
or serial). As a humidity reference you can use, for example, the Vaisala
Humidity Calibrator HMK15.
Refer to chapter Technical Data on page 57 for accuracy specifications of
the probes.
1. Connect the power/signal cable to the probe, but do not power it up
yet.
2. Place the probe in the reference environment. For example, you can
use a NaCl salt chamber (75 %RH) as the humidity reference.
3. Wait for 20 – 40 minutes for the reading to stabilize.
4. Power up the probe and wait for one minute.
5. Check the measurement reading, and compare it with the reading
that the reference should produce.