Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 39
Set Measurement Units
Use the UNIT command to view or set the measurement units that are
used in the serial line measurement messages.
UNIT [M/N]<cr>
M is for metric units, N is for non-metric units.
Quantity Metric Unit Non-Metric Unit
RH %RH %RH
Td ºC ºF
T ºC ºF
Examples:
unit
Units : Metric
unit n
Units : Non metric
Calibration Commands
Calibrate Humidity Measurement
Use the CRH command to perform a one-point or two-point correction to
the capacitance measurement of the probe. This command changes the
offset and/or gain of the measurement, depending on the calibration and
reference:
- one-point calibration with a single < 50 %RH reference will adjust
the offset of the capacitance measurement
- one-point calibration with a single > 50 %RH reference will adjust the
gain of the capacitance measurement
- two-point calibration will adjust both offset and gain. The first point
requires a < 50 %RH humidity reference, the second point must be
> 50 %RH. There must also be at least 30 percentage point difference
between the references.
CRH [reference]<cr>
When performing a one-point calibration, you need to place the probe in
the reference humidity and wait for 20 – 40 minutes for the humidity to
stabilize. To apply the adjustment, enter the CRH command with the
reference %RH as a parameter.