Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 25
CHAPTER 4
OPERATION
This chapter contains information that is needed to operate HMP60 and
HMP110.
Getting Started
When the probe has been connected to a power supply, there is a short
start-up delay before the measurement data is available through the
selected output(s) - analog or RS-485. The delay depends on the
operating voltage that is supplied to the probe:
- Operating voltage 5 ... 14 V: one second start-up delay
- Operating voltage 14 ... 28 V: four second start-up delay
Serial Line Communication
HMP60 and HMP110 support two-wire RS-485 communication. The
RS-485 interface is non-isolated and offers a maximum communications
rate of 57600 bits/s.
There is no internal termination for the RS-485 on the probe. Use of
termination resistors is not recommended. If the resistors are used, the
possible increase in current consumption should be taken into account.
Connecting to the Serial Interface
The connection to the serial interface is via the 4-pin connector on the
probe; see connector pinout on page 20.
For temporary use of the serial interface (for example, calibration), you
can use the optional USB cable (Vaisala order code: 219690). Before you
can use the USB cable, you must install the provided USB driver on your
PC, see Installing the Driver for the USB Cable on page 26.