Przeglądanie stron 77
HMP260 SERIES
HMP260-U017en-1.5 Appendix 1: Serial commands
FORM Setting the output format
FORM <cr>
"xxx...xxx"
? zzz...zzz <cr>
xxx...xxx = old format
zzz...zzz = new format
The FORM command sets the format of the outputs generated in RUN mode and by
SEND command. Please note that capital and small letters have different
meanings.
\UU..UU\ relative humidity
\TT..TT\ temperature
\DD..DD\ dewpoint temperature
\AA..AA\ absolute humidity
\XX..XX\ mixing ratio
\WW..WW\ wet bulb temperature
\uu..uu\ unit according to the preceding variable
\n line feed <lf>
\r carriage return <cr>
\t horizontal tabulation <ht> or <tab>
\\ \
For example:
format: output:
\UUU.UU\ \+TT.TT\\r 100.00 +99.99 <cr>
\TTT.T\ \uu\\r\n 15.2 'C <cr><lf>
\UUU.U\ \uuu\\+DD.D\ \uu\\r 46.9 %RH +10.8 'C <cr>
Any text can be written in the command and it appears in the output. For example:
RH: \UUU.U\ T: \+TT.TT\\r RH: 54.0 T: +25 <cr>
The format can be deleted by giving \ as a parameter:
FORM \<cr> Note. only one space before \ and none after