HMP260 SERIES
Appendix 1: Serial commands HMP260-U017en-1.5
ASEL RH T <cr>
Ch1 (RH) lo 0.000 %RH ? <cr>
Ch1 (RH) hi 100.000 %RH ? <cr>
Ch2 (T ) lo -40.000 'C ? 0 <cr>
Ch2 (T ) hi +160.000 'C ? 100 <cr>
When the ASEL command is given on its own, the transmitter outputs its current
settings:
ASEL <cr>
Ch1 (RH) lo 0.000 %RH
Ch1 (RH) hi 100.000 %RH
Ch2 (T ) lo -40.000 'C
Ch2 (T ) hi +160.000 'C
The outputs and their scales can also be given directly with the ASEL command.
ASEL xxx yyy aaa.a bbb.b ccc.c ddd.d <cr>
xxx = channel 1's quantity
yyy = channel 2's quantity
aaa.a = lower limit of channel 1
bbb.b = upper limit of channel 1
ccc.c = lower limit of channel 2
ddd.d = upper limit of channel 2
ASCL Scaling the analogue outputs
Disconnect the security lock jumper!
ASCL <cr>
Scales the outputs selected on channels 1 and 2.
For example, scaling relative humidity on the range of 0...100 %RH and temperature -
40...+160 °C:
ASCL <cr>
Ch1 (RH) lo 0.000 %RH ? <cr>
Ch1 (RH) hi 100.000 %RH ? <cr>
Ch2 (T ) lo 0.000 'C ? -40 <cr>
Ch2 (T ) hi 100.000 'C ? 160 <cr>