HMP260 SERIES
Operating Manual HMP260-U017en-1.5
34 1997-11-25
5.4.2 Output via the serial bus
5.4.2.1 Starting the measurement output
R <cr>
Starts output of measurements to the peripheral devices (RUN mode); the only
available command is S (stop).
The output mode can be changed with command FORM (see Appendix 1).
5.4.2.2 Stopping the measurement output
S<cr>
Ends the RUN mode; after this command all other commands are available.
5.4.2.3 Outputting the reading once
SEND <cr> in STOP mode
or
SEND aa <cr> in POLL mode
aa = address of the transmitter when more than one transmitter is
connected to a serial bus (0...99)
The output format depends on which parameters the transmitter can output.
Output types:
"RH=999.9 %RH T=999.9 *C",<cr><lf>
"RH=999.9 %RH T=999.9 *C Td=9999.9 *C",<cr><lf>
"RH=999.9 %RH T=999.9 *C a=9999.9 g/m3 x=9999.9 g/kg Tw=999.9*C",<cr><lf>
"RH=999.9 %RH T=999.9 *C Td=9999.9 *C a=9999.9 g/m3 x=9999.9 g/kg Tw=999.9 *C",<cr><lf>
The output mode can be changed with command FORM (see Appendix 1).