Vaisala HMP260 SERIES Instrukcja Obsługi Strona 107

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 118
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 106
HMP260 SERIES
HMP260-U017en-1.5 Appendix 4: Digital current loop module
Dual loop wiring
Set the address of the transmitter, it can be any number between 1 and 99. In
this example the address is 22:
>addr 22
Address : 22
Set the serial bus settings according to your system. This setting will become
valid after next RESET or power off:
>seri 2400 e 7 1 f
2400 E 7 1 FDX
Switch echo on:
>echo on
ECHO : ON
>
Change the serial output mode into POLL:
>smode poll
Serial mode : POLL
NOTE 1
The SMODE command must be given last.
NOTE 2
The transmitter outputs no prompt (>) after the SMODE
POLL command and it only reacts to commands which include
its address.
Check that the transmitter responds to its address:
>send 22
RH= 24.4 %RH T= 29.1 'C
Disconnect the terminal.
Remount the digital current loop module.
Close the cover.
Repeat this setting procedure with each transmitter
When all transmitters on the network have been configured, switch them off.
Przeglądanie stron 106
1 2 ... 102 103 104 105 106 107 108 109 110 111 112 ... 117 118

Komentarze do niniejszej Instrukcji

Brak uwag