output range from 500 to 1100 hPa were done. The programmable logic control was
set up to steps of 50 hPa within a timeframe of 15 minutes.
Pressure Measurement as a Function of Time
The graph of pressure as a function of time gives an account of the excellent accuracy
of the digital barometers from Vaisala.
Figure 5.7: Pressure Measurement as a Function of Time
There are almost no variances in pressure readings, neither according to the climate
chamber reference device DPI 740 nor to the sensors itself.
The overshooting reflects the process of line-up.