Vaisala HUMICAP HMT330 SERIES Podręcznik Użytkownika Strona 40

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 178
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 39
User's Guide ______________________________________________________________________
40 __________________________________________________________________ M210566EN-H
Probe Mounting
In humidity measurement and especially in calibration it is essential that
temperature of the probe and measuring environment is the same. Even a
small difference in temperature between the environment and the probe
causes an error. As the curve below shows, if the temperature is +20 °C
and the relative humidity 100 %RH, a difference of ±1 °C between the
environment and the probe causes an error of ±6 %RH.
The graph below illustrates the measurement error at 100 %RH when the
difference between the ambient and sensor temperature is 1 °C.
0507-023
Figure 22 Measurement Error at 100 %RH
Przeglądanie stron 39
1 2 ... 35 36 37 38 39 40 41 42 43 44 45 ... 177 178

Komentarze do niniejszej Instrukcji

Brak uwag