User's Guide ______________________________________________________________________
152 _________________________________________________________________ M210566EN-H
Inputs and Outputs
Operating voltage 10 ... 35 VDC, 24 VAC
with optional power supply module 100 ... 240 VAC, 50/60 Hz
Start-up time after power-up 3 s
Power consumption @ 20 °C (Uin 24VDC)
RS-232 max 25 mA
Uout 2 × 0 ... 1V / 0 ... 5V / 0 ... 10V max 25 mA
Iout 2 × 0 ... 20 mA max 60 mA
display and backlight + 20 mA
during sensor purge + 110 mA max
Analog outputs (2 standard, 3rd optional)
current output 0 ... 20 mA, 4 ... 20 mA
voltage output 0 ... 1 V, 0 ... 5 V, 0 ... 10 V
Accuracy of analog outputs at 20 °C ± 0.05 % full scale
Temperature dependence of the analog outputs ± 0.005 %/°C full scale
External loads
current outputs RL < 500 ohm
0 ... 1V output RL > 2 kohm
0 ... 5V and 0 ... 10V outputs RL > 10 kohm
Max wire size 0.5 mm2 (AWG 20)
stranded wires recommended
Digital outputs RS-232, RS-485 (optional)
Relay outputs (optional) 0.5 A, 250 VAC, SPDT
Display (optional) LCD with backlight, graphic
trend display
Menu languages English, French, Spanish,
German, Japanese, Russian,
Swedish, Finnish, Chinese.
Mechanics
Cable bushing M20x1.5 for cable diameter
8 ... 11mm/0.31..0.43"
Conduit fitting 1/2"NPT
User cable connector (optional) M12 series 8- pin (male)
option 1 with plug (female) with 5 m /
16.4 ft black cable
option 2 with plug (female) with screw
terminals
Probe cable diameter
HMT333 80ºC 6.0 mm
Other probes 5.5 mm
Probe cable lengths 2 m, 5 m or 10 m
Probe tube material
HMT331 Chromed ABS plastic
Other probes AISI 316L
Housing material G-AlSi 10 Mg (DIN 1725)
Housing classification IP 65 (NEMA 4)