User's Guide ______________________________________________________________________
84 __________________________________________________________________ M210855EN-C
Address : 0
Echo : ON
Module 1 : BARO-1
Module 2 : BARO-1
Module 3 : BARO-1
Module 4 : EMPTY
>
ERRS
Use the command ERRS to display all unacknowledged errors (and
clear them). If an error is still active, it will be displayed again with the
ERRS command.
Example:
>errs
PASS
No errors
>errs
FAIL
Error: Operating voltage out of range.
>
VERS
Use the command VERS to display software version information.
Example:
>vers
PTB330 / 1.00
>
Resetting barometer By Using
Serial Line
RESET
Resets the device. The user port switches to start-up output mode
selected with command SMODE.