Chapter 3 ________________________________________________________________ Installation
VAISALA ________________________________________________________________________ 45
0507-024
Figure 25 Horizontal Mounting of Probe
The following numbers refer to Figure 25 above:
1
=
To be sealed
2
=
To be insulated
3
=
Insulate the cable
4
=
Let the cable hang loosely. This prevents condensed water
running to the sensor along the cable.
0507-022
Figure 26 Vertical Mounting of Probe
The following numbers refer to Figure 26 above:
1
=
To be sealed
2
=
Insulate the cable
3
=
To be insulated
4
=
Let the cable hang loosely. This prevents condensed water
running to the sensor along the cable.