Non-metric units are not available on Modbus. If they are needed,
calculate the conversion outside the transmitter.
Configuration flags are used to select some basic options of the
instrument and to manually start the sensor operations.
Table 48 Configuration Flag Registers
Standard filtering on/off
Extended filtering on/off
1 = Extended filtering on
Automatic chemical purge on/off
See configuration register
Start-up chemical purge on/off
Chemical purge in progress
1 = Purge in progress
(write 1 for manual start)
1 = Sensor heating on
(optional feature)
Flags 1281 and 1282 are mutually exclusive.
The pressure compensation value is set with configuration registers
0769…0770 (1025) or 0771…0772 (1026).
Some of the configuration registers might not have an effect depending
on the instrument model and configuration.
Other configuration settings are not usually needed and are not available
using the Modbus interface. The service cable must be used to change
other configuration settings if needed.
Exception Status Outputs
Exception status outputs (read with function code 07) tell the summary of
the PTU300 transmitter status as described in Table 49 below.
Table 49 PTU300 Exception Status Outputs
1 = Online data available
Status information is available also with register access; see section
Status Registers (Read-Only) on page 211.