Vaisala PTU300 Podręcznik Użytkownika Strona 176

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 217
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 175
User's Guide _______________________________________________________________________
174 __________________________________________________________________ M210796EN-G
Relative Humidity Adjustment After Sensor
Change
Using Display/Keypad
When using the optional display/keypad, follow the instructions in
section Using Display/Keypad on page 168 on page but select Adj. for
new RH sensor (instead of 1-point/ 2-point adjustment).
Using Serial Line
After sensor change, carry out the procedure as described in previous
sections. Just replace the CRH command with the FCRH command.
FCRH
Example:
>fcrh
RH : 1.82 1. ref ? 0
Press any key when ready...
RH : 74.22 2. ref ? 75
OK
>
The OK indicates that the calibration has succeeded.
Przeglądanie stron 175
1 2 ... 171 172 173 174 175 176 177 178 179 180 181 ... 216 217

Komentarze do niniejszej Instrukcji

Brak uwag