WS425 F/G ASOS Profile
Configurable Parameters
Table 43 below lists the configurable parameters and their
allowed and default values for the WS425 F/G ASOS profile.
Table 43 Configurable Parameters for WS425 F/G ASOS
Profile
Profile-Specific Allowed
Values
disabled
Automatic data message number.
Selects the data message format for
Profile for serial port COM1
Profile for serial port COM2
1 = RS-422
Interface for serial port COM2
0.25 … 3600
Resolution: 0.25
Averaging time for wind
measurement in seconds
0.25 … 3600
Resolution: 0.25
Averaging time for wind minimum and
maximum in seconds
0 = Array facing up
1 = Array facing down
Orientation of the array of WMT700
0 = Meters per second (m/s)
1 = Miles per hour (mph)
2 = Kilometers per hour (km/h)
* The parameter has no protocol-specific default value. It is determined
separately in the configuration code.
**
Wind speed unit and COM2 digital communication interface are defined in
the order form. Default setting may be any of the allowed options, and it can be
verified from the unit’s configuration code.
To start continuous measurement, use the START command.
You can stop the measurement with the STOP command. For
instructions, refer to section Wind Measurement Control
Commands on page 115.