Appendix A _________________________________________________________ Modbus Reference
VAISALA ________________________________________________________________________ 71
Modbus Register Map
Table 14 Modbus Register Map
Logical
Address
(decimal)
PDU Address
(hexadecimal)
Register Description
Data
Format
Register
Type
Notes
3
00 02
T
(measured)
LSW
MSW
32-bit
float
read-only
°C
4
00 03
29
00 1C
a
w
(measured)
LSW
MSW
32-bit
float
read-only
30
00 1D
35
00 22
H
2
O (oil) (measured) *
LSW
MSW
32-bit
float
read-only
ppm
36
00 23
513
02 00
Fault status
16-bit
boolean
read-only
1 = no errors
516
02 03
Error Code
LSW
MSW
32-bit
bit field
read-only
0 = no errors
517
02 04
785
03 10
Oil factor A (set value) *
LSW
MSW
32-bit
float
read/write
786
03 11
787
03 12
Oil factor B (set value) *
LSW
MSW
32-bit
float
read/write
788
03 13
1537
06 00
Address
16-bit
integer
read/write
1…255
(1247 allowed
in Modbus)
* Availability of the register depends on the device model.
PDU address
Actual address bytes used in a Modbus Protocol Data unit.
LSW
Least significant word (bits 15 … 0).
MSW
Most significant word (bits 31 … 16).
16-bit integer
Numeric value in range 0 … 65535.
16-bit boolean
Numeric value, 0 or 1.
32-bit bit field
32 individual values, each 0 or 1.
32-bit float
Floating point number encoded according to IEEE 754.
read-only
Register value cannot be changed with Modbus functions.
read/write
Register value can be changed with Modbus functions.