Set Voltage for LED Alarm
With the LED command, you can set a different voltage level for each
channel.
LED<cr>
Example:
led ?
Ch 1 Led Voltage : 2.80 V
Ch 2 Led Voltage : 2.80 V
Extend Analog Output Range
Use the AOVER command to allow the analog output channels to
exceed their specified range by 10 %. The scaling of the parameter
remains as before; the extra range is used for additional measurement
range in the wet end.
AOVER [ON/OFF]<cr>
Example:
aover on
: ON
Serial Line Output Commands
Start Measurement Output
Use the R command to start the continuous outputting of measurement
values as an ASCII text string to the serial line. The format of the
measurement message is set with the FORM command.
R<cr>
Example:
r
aw= 0.261 T= 23.8 'C H2O= 15 ppm
You can change the format of the output with the following commands:
- Outputting interval can be changed with the INTV command.
- Output message format can be changed with the FORM command.