15. Wait for the measured value to stabilize and press Ready. Enter the
correct value of the reference point. Press OK when done, and
YES when asked if you really want to adjust.
16. You will see the text Adjustment done. You have now completed
the 2-point adjustment. Select Back to return to the parameter
selection screen, and Exit to leave the adjustment mode.
17. Remove the MMT162 from the reference environment and reattach
the filter.