Vaisala HUMICAP HMW90 SERIES Dokumentacja Strona 59

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 74
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 58
553
HUMIDITY
HUMIDITY
9
WE MAKE IT EASY.
kele.com 888-397-5353 USA
March 2014
Space (Wall) MountCover (all mounts)
Duct Mount
1.28
(3.24)
Cover Base
Wiring
Access
4.70
(11.92)
4.70
(11.92)
3.80
(9.63)
2.37
(6.03)
1/2''
(1.27)
knockout
(each side)
Outside
Air
Mount
4.70
(11.92)
6.38
(16.2)
2.75
(7.0)
8.80 (22.37)
4.70
(11.92)
0.5
(1.27)
2.75
(7.0)
in
(cm)
DIMENSIONS
DEW POINT, ENTHALPY & WET BULB TRANSMITTERS
DP4A, EN4A, WB4A
SW3
SW1
VI
VI
J8
TH
TH
I2-TMP
I2XR/P–
I1XR/P+Vin
XR RTN
TEMP RTN
Switch Positions for Voltage Mode Switch Positions for Current Mode
VOLTAGE MODE
SW3
SW1
VI
VI
XR and Temperature - Voltage
CURRENT MODE
XR and Temperature - Current Loop
Switch Selectable 0-5 V or 0-10 V Output
I1DP/P+
I2DP/P–
V-XR
GND
V-TMP
J7
10V
SW2
5V
24 VDC
+
XR
Temp
J7
J8
Voltage
Current
JPR1
3K
10K
10V
SW2
5V
SW1
VI
SW3
VI
Connector and Switch Locations
+
+
Not
used
Only one terminal block provided; move for
either voltage or current.
*XR represents; DP - Dewpoint;
EN-Enthalpy; WB - Wet bulb
WIRING
ORDERING INFORMATION
MODEL DESCRIPTION
EN4A-D Duct enthalpy and dry bulb transmitter
EN4A-O OSA enthalpy and dry bulb transmitter
DP4A-D Duct dew point and dry bulb transmitter
EN4A-S Room enthalpy and dry bulb transmitter
DP4A-O OSA/Wall dew point and dry bulb transmitter
DP4A-S Room dew point and dry bulb transmitter
WB4A-D Duct wet bulb and dry bulb transmitter
WB4A-O OSA wet bulb and dry bulb transmitter
WB4A-S Room wet bulb and dry bulb transmitter
RELATED PRODUCTS PAGE
DCPA-1.2 Power supply, 120 VAC IN to 24 VAC/24 VDC OUT 994
DCP-1.5-W Power supply, 24 VAC IN to 24 VDC OUT 995
Przeglądanie stron 58
1 2 ... 54 55 56 57 58 59 60 61 62 63 64 ... 73 74

Komentarze do niniejszej Instrukcji

Brak uwag